Maintenance of direct reading spectrometer


I. Maintenance of Stimulator

The basic maintenance part of the excitation system is inside the excitation table. Whether the inside of the excitation table is clean, whether the electrode pole distance is stable, and the height of the arc arc flame of the excitation table relative to the optical system will affect our data results.

1. Pull out the exhaust pipe and use a vacuum cleaner to clean up the dust in the excitation cavity.

2.Wipe the sheath with absorbent cotton and pure alcohol

3. After the cleaning is completed, re-measure the analysis interval with a polar gauge.

4. Rinse the entire circuit with argon for 2 ~ 3 minutes to flush out the air entering the spark table.

Second, the maintenance of the lens

The condenser lens is the only way for the light emitted by the sample to enter the spectrometer. The mirror must be kept clean. However, the long-term use of the condenser will be contaminated, and a layer of dark brown substance is attached to its outer surface. The longer the time, the thicker the deposit. The metal powder generated by the excitation will also adhere to the outer surface of the lens, reducing the light transmittance and reducing the absolute light intensity of all elements, especially the far ultraviolet elements C, P, S are more obvious.

The surface of the condenser lens is made of absorbent cotton, dipped in a mixture of ether and ethanol, and wiped from the center of the condenser outwards in a circular motion. If the pollution is serious, you can wipe it several times with more force until you wipe off the dirt and then wipe the watermark with a dry cotton ball.

Special attention should be paid to the fact that the condenser lens is made of quartz. Do not use hard objects (never use cotton tweezers to clamp cotton) to avoid scratching the lens surface.

Third, the maintenance of waste gas barrels

Argon is an inert gas and has no direct harm to the human body. However, part of the dust generated by the spark table after the sample is excited will be blown away by argon, and the exhaust pipe will become black. After a long time, it needs to be cleaned to ensure the smoothness and cleanliness of the gas path. The main component of the exhaust gas generated after the excitation is that argon contains the excited metal dust. After long-term direct inhalation, it will cause silicosis and eye damage, which will cause great harm to the human body. The exhaust pipe of the direct-reading spectrometer can be a transparent plastic pipe, which can be inserted into the water during installation; if the pipe wall becomes black and opaque, the exhaust pipe needs to be replaced.


Nanjing Kirin Scientific Instrument Group Co., Ltd.

Testing Center

November 29, 2019


relevant information

    麒麟

    Over 20 years

    Focus on elemental analysis technology development

    麒麟

    Testing Equipment

    Exported to many countries in Southeast Asia

    麒麟

    More than 10,000 companies

    Trust and praise

    麒麟

    48 detections

    Quality assurance

    麒麟

    24 hours

    All-day service model

    AG真人网 电玩竞技游戏 皇冠体育官网 真人斗牛app 皇冠体育app 3d捕鱼 电玩游戏平台 电玩竞技游戏